Sign In | Join Free | My polstate.com |
|
Description : Continuous Emission Monitoring System for HCl and CO in Waste Incineration Flue Gas (CEMS)
Stock : 1set
MOQ : 1set
Price : USD26000
Payment Terms : T/T
Shipping Method : LCL
product name : online gas analyzer
Range : 0-100 Ppm/Mg/M3
Response Time : Less Than 120s
material : Alloy, Metal, Stainless Steel
Power Supply : 220V ±10%, 50Hz ±1Hz, 5 kVA
Light Path : 760mm
WT Online Gas Monitoring System
1. Basic Design Principle of the WT-400 Analysis System
The WT-400 Ammonia Analysis System adopts a high-temperature extractive monitoring method, with a laser-based NH₃ analyzer as the core unit. All gas-contacting components are heated and insulated to above 180℃ throughout the sampling path, ensuring stable and accurate NH₃ measurements even under challenging process conditions.
2. Operating Conditions and Requirements
2.1 General Process Conditions
Maximum dust concentration: < 2000 g/Nm³
Flue gas temperature at the sampling point: ≤ 600℃
Sampling point pressure: -15 to 15 MPa
Tar and benzene content: < 10 g/Nm³
Presence of saturated water vapor
2.2 Environmental Conditions
Temperature control cabinet installed outdoors; analyzer unit installed indoors
Ambient temperature: 5–45℃
Relative humidity: 20–85% RH
Atmospheric pressure: 10–106 kPa (depending on site location)
2.3 Utilities and Gas Supply
Power Supply: 220V ±10%, 50Hz ±1Hz; Power consumption: 5 kVA; free from strong electrical interference
Compressed Air: For purging the sampling system to prevent clogging; must be clean, dry, oil-free, with a pressure of 4–7 kg/cm² to ensure effective cleaning
Standard Gas: For system preprocessing, calibration, and performance verification; must meet the specified purity requirements to maintain measurement accuracy
No. | Standard Gas Compositions | Standard Gas Concentration | Standard Gas Pressure |
1 | N2 | ≥ 99.9999% | 0.5 -10 MPa |
2 | NH3 | 80% of the measurement rage | 0.5 -10 MPa |
3. Features of the WTKF200 Ammonia Escape Online Monitoring System
Integrated Sampling Probe – Compact, streamlined design with large dust holding capacity; supports dust filtration up to ≤ 100 g/Nm³ without clogging.
High Filtration Precision – Dust filtration accuracy < 0.1 μm.
High Measurement Accuracy – Analysis accuracy of ±1% FS.
Fast Response – Response time (T90) ≤ 30 s.
High Reliability – Mean Time Between Failures (MTBF) > 3 years.
Low Maintenance – Maintenance interval exceeds 1 year.
Distortion-Free Sampling – High-temperature extractive sampling method ensures accurate, representative measurements.
Temperature-Controlled Sampling – High-temperature, constant-temperature design maintains NH₃ measurement accuracy.
Automatic Purging – Built-in automatic cleaning system removes dust and prevents blockages.
Signal Output – 4–20 mA output; control alarm signal NO/NC, 1A/220V.
We offer fully customized gas online monitoring solutions tailored to the customer’s unique requirements and operating environments, ensuring optimal performance and measurement accuracy.
![]() |
Continuous Emission Monitoring System for Flue Gas HCl Co Waste Incineration CEMS Gas Analyzers Images |